New Approach Objectives:

  1. Avoid Orthogonality Issues
  2. Allow for more sensors to be used
  3. Make Calibrations more accurate
  4. Speed up Calibration Process
  5. Improve MWD Accuracy

New Approach

  • For Example: T W
    • X is 0 90
    • Y is 90 90
    • Z is 0 0
  • Microtesla 4AM :T W
    • A is -30 25
    • B is 30 90
    • C is -90 90
    • D is 110 150
  • Find the 2 alignments for each sensor
    • Alignment from along hole axis W
    • Alignment from high side axis T
  • Include all sensors available in the tool
  • Calibrate each sensor’s 2 alignments v Temperature using a ‘hot’ tumble and a ‘cold’ tumble (Use a linear interpolation for W and T at other temperatures)
  • Using actual alignments corrected at each temperature
  • Set tool to inc 60, azi 0, Tf -45, heat to 150 and record while cooling
  • Set tool to inc 120, azi 180, Tf 135, heat to 150 and record while cooling
  • Determine least squares best fit polynomials to correct for scale and bias
  • In Instrument firmware use temperature corrected sensor data and actual alignments to produce synthetic 3 axis perfectly orthogonal data for surveys

....

View the entire Presentation:

MWD: A New approach

 